产品名称: 530L流量校准器Bios Defender 530-Low Flow
产品型号: 530L
产品厂商: 上海豫东电子科技有限公司
产品文档: 无相关文档

简单介绍
530L流量校准器Bios Defender 530-Low Flow ,Bios Defender 530 Standardized & Volumetric FlowPrimary Standard Air Flow Calibrators Air Flow Calibrators with DryCal? Technology The Bios Defender 530

530L流量校准器Bios Defender 530-Low Flow

的详细介绍

流量校准器Defender 510/520/530 仪器介绍:

    Defender 510/520/530是新一代的干式气体流量计,它是一级流量计,并被众多的个体采样泵厂商推荐和配套使用。可用于工业卫生、环保领域和各类实验室。它采用干式活塞运动的原理,一次按键即可产生连续多个读数。与传统的泡沫式流量计相比,使用更加方便快速。520型号比510增加了同时测试温度和压力的功能。

流量校准器Defender 510/520/530  主要特点:

  • 采用干式活塞原理,使用简单方便
  • 流量精确控制,经过ISO17025认证
  • 结构紧凑
  • 宽大显示屏幕
  • 每组读数高达100个
  • 可内置温度、压力传感器,优化测试结果(520系列)
技术参数:

型号
Defender 510
Defender 520
Defender 530
流量范围
510L, 5-500ml/min
510M, 50-5000ml/min
510H, 300-30,000ml/min
520L, 5-500ml/min
520M, 50-5000ml/min
520H, 300-30,000ml/min
530L, 5-500ml/min
530M, 50-5000ml/min
530H, 300-30,000ml/min
流量精度
1%
1%
1%
标准的精度
N/A
N/A
1.2%
可选
测量
单位
流量
mL/min, L/min,
cc/min, cf/min
mL/min, L/min,
cc/min, cf/min
mL/min, L/min,
cc/min, cf/min
标准化
N/A
N/A
smL/min, sL/min,
scc/min, scf/min
温度
N/A
℃ ℉
℃ ℉
压力
N/A
mmHg, PSI, kPa
mmHg, PSI, kPa
温度和压力传感器
No
Yes
Yes
压力
N/A
3.5 mmHg (典型)
7.0 mm (最大)
3.5 mmHg (典型)
7.0 mm (最大)
温度
N/A
0.8℃ (典型)
1.3℃ (最大)
0.8℃ (典型)
1.3℃ (最大)
每次读数时间
1~15 秒
读数模式 (选择从1到100次测量的平均值)
单次 (手动), 连续或突发模式
气体兼容性
吴腐蚀性气体、不结露气体, 湿度小于70%
流动模式
压迫或抽吸
测量单元
一体式
AC 交流直流变换器/充电器
12V DC, > 250ma, 2.5mm中心位置
Battery System电池类型
6V 可充电密封铅酸蓄电池,典型工作时间6~8 小时
电池运行时间 (5 次/分钟)
背景光打开3小时, 背景光关闭 8小时
压入或抽吸附件
软管
操作压力 (绝对)
15 PSI
显示
带背光的图形LCD液晶显示器
数据端口 (使用优化软件)
RS-232串口
数据电缆 (使用优化软件)
1 米 Defender校准器至电脑的串口
防护箱
软质或硬质仪器箱
尺寸(H x W x D)
5.5”x6”x3”/140x150x75mm
重量
29oz/820g
服从:RoHS-和CE


530L流量校准器Bios Defender 530-Low Flow

 

Bios Defender 530 Calibrator -- Coming January 2010

The Bios Defender 530, with the ability to take both volumetric and standardized readings, is the latest addition to the popular Defender 510 and 520 line of Bios DryCal primary flow calibrators. Automatic correction of readings for temperature and pressure, rather than merely recording of ambient conditions, is essential for many applications.

The Defender 530, like the other instruments in the line, provides users with a true primary flow calibrator for industrial hygiene, environmental and laboratory applications. The design offers low cost, light weight and ultra-compact size, incorporating the highly-regarded DryCal technology. It is small enough to fit easily into a slim briefcase for convenient transport.

Like all Bios calibrators, the Defender 530 is easy to use.  Turn it on. Connect it to a sampling pump. Start taking readings. It's as simple as that. Calibrations within a minute with no messy soap bubble solutions.

The Defender 510, 520 and 530 can take hands-free continuous readings while quickly and accurately setting the flow rate on air sampling pumps or other instruments. The Defender 510/520/530 calibrators connect to any PC and pre- and post-calibration data can be downloaded with the included Integrator 110 with the Optimizer Collect Light software. 


530L流量校准器Bios Defender 530-Low Flow


Bios Defender 530 Specifications

530L流量校准器Bios Defender 530-Low Flow

Size5.5"H x 6"W x 3"D
140 mm x 150 mm x 75 mm
Weight29 oz. / 820 g
Flow Ranges 
   DryCal Defender Low (L) Flow:5 to 500 ccm
   DryCal  Defender Medium (M) Flow:50 - 5,000 ccm 
   DryCal Defender High (H) Flow:300 - 30,000 ccm
Accuracy 
   Volumetric ±1.0%
   Standardized±1.2%
User-Selectable Measurement Units
   Volumetric FlowmL/min, L/min, cc/min, cf/min
   Standarized FlowsmL/min, sL/min, scc/min, scf/min
   Pressure mmHg, PSI, kPa
   Temperature ºC, ºF
Readings 
   Approximate Time per Reading1-15 seconds
   Reading StylesSingle (manual), Continuous or Burst, with averaging function user-selectable from 1 to 100 measurements
Gas CompatibilityUse with non-corrosive, non-condensing, non-combustible gases, less than 70% humidity
Flow ModesSuction or Pressure
Measurement Cell StyleIntegrated
Temperature & Pressure SensorsIn the flow stream
Press: 3.5 mmHg (typical), 7.0 mm (max)
Temp: 0.8 °C (typical) 1.3 °C (max)
Battery System 
   AC Adapter/Charger12 DC, >250 ma, 2.5 mm, center positive
   Battery 6V rechargeable, sealed lead-acid, 6-8 hours typical operation
   Battery Operational Time (5 cycles/min), 3 hours backlight on, 8 hours backlight off
Pressure & Suction Fittings¼” ID Hose Barb
DisplayBacklit graphical LCD
Storage Temperature0-70 °C
Ambient Humidity0-70% - non-condensing
Operating Pressure (Absolute) 15 PSI
WarrantyInstrument: 1 year, Battery: 6 months
Data Port
Serial RS-232 (for use with Optimizer Collect software)
Data Cable1 meter (Defender Data Port to PC Serial Port)
Protective CaseSoft side or Pelican cases available
Laboratory AccreditationBacked by ISO 17025 accreditation by NVLAP of NIST
Instrument CertificationRoHS and CE Compliant

Bios Defender 500 Kits Include:  One flow cell, Bios Optimizer 110 PC Serial Cable, Software CD, AC Power Adapter/Charger, Plastic Leak Test Caps (2), NIST-Traceable Calibration Certificate and Mini-Manual.

Bios Integrator 110 and Optimizer Collect Software

The Defender 510/520 and the Defender 530 come standard with the Bios Integrator 110 and Optimizer Collect Light software for bullet-proof transfer of an ASCII text file of the flow data into an Excel® spread sheet with automatic recording and graphing of flow measurements.

The Optimizer Collect software helps to meet the needs of ISO, TQM, or occupational health and safety management systems by providing a secure audit trail for all pump calibrations, regardless of the pump type or manufacturer.

The basic Bios Optimizer Collect Light allows users to download data directly to a Microsoft Excel® preconfigured Spreadsheet.for easier collection, storage and record-keeping.

Enter flow rates from pumps or other flow source and calibrate the flow source. Data is recorded directly to a preconfigured Microsoft Excel® Spreadsheet.

 
Enter flow data from a flow meter and capture and compare the flow measurements from tje Bios DryCal precision calibrator into a preconfigured Microsoft Excel® Spreadsheet.
Optiimizer Collect light is upgradeable to a full version of Collect by purchasing an Activation code from Bios for automatic multi-point calibration of MFC/MFM using DryCal and Integrator 110.

The upgrade to the Bios Optimizer Collect full version allows the user to not only capture, but secure, organize, search and present data.  It stores the records in a small but secure SQL database. These records can also be easily opened in a variety of off-the-shelf database applications, including Microsoft Excel and Access. 

 

The Bios Defender - Overview

Accuracy

  • Proven percent of reading accuracy is supported by ISO 17025 accreditation by NVLAP of NIST
  • Integrated temperature sensors and pressure transducers placed directly in the flow stream for optimum standardized results.

Convenience

  •  Housed in a molded case with rubber handgrips for a better feel and a sleeker, more contemporary look
  •  Large illuminated flow cell viewing window
  •  Navigation menu and a large, illuminated graphical display with zoom feature
  •  Rated for extreme cold operation, down to -20 ºC
  •  RoHS (Restriction of Hazardous Substances) compliant for Europe
  •  Automatic Shutoff

Flexibility

  •  Choose between flow units
    • – Volumetric: mL/min, L/min, cc/min, or cf/min
    •   Standardized: smL/min, L/min, cc/min, scf/min
  •  Include up to one hundred readings in a group average
  •  Take individual readings, continuous hands-free readings, or set to read in time intervals anywhere from one minute to sixty minutes between readings
  •  Available in three models for application expansion. Use the
    • Bios Defender 510 to calibrate personal air samplers and other volumetric instruments;
    • Bios Defender 520 "three-in-one" to calibrate volumetric instruments, record ambient temperature and pressure conditions for the audit trail, or to calibrate devices that read in mass flow;
    • Bios Defender 530, for both volumetric flow readings or standarized readings, which correct for temperature and pressure, or to calibrate devices that read in mass flow,.

Software:

  • The Integrator 110 and Optimizer Collect Light software is bundled with the Defender with an integrated serial PC interface for easy transfer of flow data to a text file, with automatic recording and graphing of flow measurements.
  • The upgrade to the Bios Optimizer Collect Full version allows the user to not only capture, but secure, organize, search and present data. It stores the records in a small but secure SQL database. These records can also be easily opened in a variety of off-the-shelf database applications, including Microsoft Excel and Access.