三氯乙烯检测管132L Trichloroethylene
Trichloroethylene No.132L
Cl2C:CHCl
| Performance | 
| MeasuringRange | 1to 2ppm | 2to 25ppm | 25to 70ppm | 
| Number of Pump Strokes | 2 | 1 | 1/2 | 
| Correction Factor | 1/2 | 1 | 2.8 | 
| SamplingTime | 1 minute per pumpstroke | ||
| DetectingLimit | 0.4 ppm (n=2) | ||
| ColorChange | Yellow →Purple | ||
| ReactionPrinciple | Cl2C:CHCl + PbO2 +H2SO4 →HCl | ||
| Coefficient ofVariation | 10% (for1 to 5 ppm), 5% (for 5 to 25 ppm) | ||
| ShelfLife | 2Years | ||
| Correctionsfor temperature & humidity | Temperature correction is necessary | ||
| Store the tubes inthe refrigerator to keep at 10oC (50oF) orbelow. | |||
| Possible coexisting substances and theirinterferences | (NOTE) | 
| Substance | Concentration | Interference | Change color byitself | 
| Nitric Oxide,Nitrogen dioxide | - | No effect | Nodiscoloration | 
| Hydrogen chloride,Chlorine, Bromine | - | Plus error | Produce reddishpurple stain | 
| Acetone | 200 ppm | No effect | Nodiscoloration | 
| UnsaturatedHalogenated HCs | - | Plus error | Produces reddishpurple stain | 
| Aromatichydrocarbons | 100ppm | Minus error | Nodiscoloration | 
| Othersubstance measurable with this detector tube | 
| Substance | Correction Factor | Pump Strokes | Measuring Range | 
| Benzylchloride | 0.8 | 2 | 1.6 to 20ppm | 
| Calibration gasgeneration | Diffusion tubemethod | |
| TLV-TWA | TLV-STEL | Explosive range | 
| 50ppm | 100ppm | - | 
 
    