丙酮检测管151L Acetone CH3COCH3,Measuring Range 50 to 4000 ppm 4000 to 12000 ppm Number of Pump Strokes 2 1 Correction Factor 1 3 Sampling Time 2 minutes per pump stroke Detecting Limit 5ppm(n=2)丙酮气体检测管价格
详情介绍:
丙酮检测管151L Acetone CH3COCH3
Acetone                        No.151L
CH3COCH3
丙酮检测管151L Acetone CH3COCH3
Performance

 

MeasuringRange

50 to 4000 ppm

4000 to 12000ppm

Number of Pump Strokes

2

1

Correction Factor

1

3

SamplingTime

2 minutes per pumpstroke

DetectingLimit

5 ppm (n=2)

ColorChange

Yellow → Red

ReactionPrinciple

CH3COCH3 + (NH2OH)3 • H3PO4 → H3PO4+ Base
→ H3PO4+Base

Coefficient ofVariation15% (for50 to 500 ppm), 10% (for 500 to 4000 ppm)
ShelfLife2Years
Correctionsfor
temperature & humidity
Temperature correction is necessary
Store the tubes inthe refrigerator to keep at 10oC (50oF) orbelow.



Possible coexisting substances and theirinterferences(NOTE)


Substance

Concentration

Interference

Change color by itself

Acrolein,Acetaldehyde

1/10 time

Plus error

Produce reddishstain

Aromatichydrocarbons

-

No effect

Nodiscoloration

Methyl isobutylketone

-

Plus error

Produce reddishstain

Methyl ethylketone

-

Plus error

Produce reddishstain



Othersubstance measurable with this detector tube


Substance

Correction Factor

Pump Strokes

Measuring Range

Methyl ethylketone

Factor: 0.42

5

21 to 1680 ppm

Propionaldehyde

Factor: 0.47

2

24 to 1880ppm



Calibration gasgenerationDiffusion tubemethod


 


TLV-TWATLV-STELExplosive range
500 ppm750ppm2.1to 13%