产品分类
PRODUCT CENTER
销售电话:时经理
18121126516
021-64835787
醋酸检测管81L Acetic acid CH3CO2H
Acetic acid No.81L

Performance |
Measuring Range | 0.125 to 0.25 ppm | 0.25 to 10 ppm | 10 to 25 ppm |
Number of Pump Strokes | 2 | 1 | 1/2 |
Correction Factor | 1/2 | 1 | 2.5 |
Sampling Time | 1.5 minutes per pump stroke | ||
Detecting Limit | 0.05 ppm (n=2) | ||
Color Change | Pink → Pale Yellow | ||
Reaction Principle | CH3CO2H + Base → Reaction product | ||
Coefficient of Variation | 10% (for 0.25 to 3 ppm), 5% (for 3 to 10 ppm) | ||
Shelf Life | 2 Years | ||
Corrections for temperature & humidity | Temperature correction is necessary | ||
Store the tubes at in the refrigerator to keep at 10ºC(50ºF) or below. |
Possible coexisting substances and their interferences | (NOTE) |
Substance | Concentration | Interference | Change color by itself |
Ammonia | Higher than 2 times | - | No |
Chlorine, sulphur dioxide, Nitrogen dioxide | + | Pale Yellow | |
Formic Acid, Acetic anhydride |
| + | Pale Yellow |
Other substance measurable with this detector tube |
Substance | Correction | No. of pump strokes | Measuring range |
Acetic anhydride | Factor: 0.6 | 1 | 0.15 to 6 ppm |
Acrylic acid | Factor: 1.8 | 1 | 0.45 to 18 ppm |
Butyric acid | Factor: 1.3 | 1 | 0.325 to 13 ppm |
Formic acid | Factor: 2.0 | 1 | 0.5 to 20 ppm |
Isovaleric acid | Factor: 1.5 | 1 | 0.38 to 15 ppm |
Methacrylic acid | Factor: 1.4 | 1 | 0.35 to 14 ppm |
Propionic acid | Factor: 1.0 | 1 | 0.25 to 10 ppm |
Valeric acid | Factor: 1.5 | 1 | 0.38 to 15 ppm |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
- | 15 ppm | - |