乙醛气体检测管92 CH3CHO
Acetaldehyde                         No.92
CH3CHO

Performance  | |||||
Measuring Range  | 5 to10 ppm  | 10 to 300 ppm  | 300 to 750 ppm  | ||
Number of Pump Strokes  | 4  | 2  | 1  | ||
Correction Factor  | 0.5  | 1  | 2.5  | ||
Sampling Time  | 2 minutes per pump stroke  | ||||
Detecting Limit  | 2 ppm (n=4)  | ||||
Colour Change  | Yellow → Red  | ||||
Reaction Principle  | 3CH3CHO + (NH2OH)3H3PO4 → H3PO4  | ||||
Coefficient of Variation  | 10% (for 10 to 100 ppm), 5% (for 100 to 300 ppm)  | ||||
Shelf Life  | 2 Years  | ||||
Corrections for temperature & humidity  | Temperature correction is necessary  | ||||
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.  | |||||
Possible coexisting substances and their interferences (NOTE)  | |||||
Substance  | Concentration  | Interference  | Change colour by itself to  | ||
Ammonia  | 
  | -  | No  | ||
Acrolein  | 3/5 times  | +  | Red  | ||
Acetone  | 2 times  | +  | Red  | ||
Methyl ethyl ketone  | 6 times  | +  | Red  | ||
MIBK  | 10 times  | +  | Red  | ||
Other substance measurable with this detector tube  | |||||
Substance  | Correction Factor  | No.of pump strokes  | Measuring Range  | ||
Diacetyl  | by scale  | 3  | 25 to 1500 ppm  | ||
Calibration gas generation Diffusion tube method  | |||||
TLV-TWA  | TLV-STEL  | Explosive range  | 
  | ||
-  | C 25 ppm  | 4 to 60%  | 
  | ||
Acetaldehyde No.92
CH3CHO

Performance  | |||||
Measuring Range  | 5 to10 ppm  | 10 to 300 ppm  | 300 to 750 ppm  | ||
Number of Pump Strokes  | 4  | 2  | 1  | ||
Correction Factor  | 0.5  | 1  | 2.5  | ||
Sampling Time  | 2 minutes per pump stroke  | ||||
Detecting Limit  | 2 ppm (n=4)  | ||||
Colour Change  | Yellow → Red  | ||||
Reaction Principle  | 3CH3CHO + (NH2OH)3H3PO4 → H3PO4  | ||||
Coefficient of Variation  | 10% (for 10 to 100 ppm), 5% (for 100 to 300 ppm)  | ||||
Shelf Life  | 2 Years  | ||||
Corrections for temperature & humidity  | Temperature correction is necessary  | ||||
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.  | |||||
Possible coexisting substances and their interferences (NOTE)  | |||||
Substance  | Concentration  | Interference  | Change colour by itself to  | ||
Ammonia  | 
  | -  | No  | ||
Acrolein  | 3/5 times  | +  | Red  | ||
Acetone  | 2 times  | +  | Red  | ||
Methyl ethyl ketone  | 6 times  | +  | Red  | ||
MIBK  | 10 times  | +  | Red  | ||
Other substance measurable with this detector tube  | |||||
Substance  | Correction Factor  | No.of pump strokes  | Measuring Range  | ||
Diacetyl  | by scale  | 3  | 25 to 1500 ppm  | ||
Calibration gas generation Diffusion tube method  | |||||
TLV-TWA  | TLV-STEL  | Explosive range  | 
  | ||
-  | C 25 ppm  | 4 to 60%  | 
  | ||