021-64835787
乙醛气体检测管92 CH3CHO
Acetaldehyde No.92
CH3CHO
Performance | |||||
Measuring Range | 5 to10 ppm | 10 to 300 ppm | 300 to 750 ppm | ||
Number of Pump Strokes | 4 | 2 | 1 | ||
Correction Factor | 0.5 | 1 | 2.5 | ||
Sampling Time | 2 minutes per pump stroke | ||||
Detecting Limit | 2 ppm (n=4) | ||||
Colour Change | Yellow → Red | ||||
Reaction Principle | 3CH3CHO + (NH2OH)3H3PO4 → H3PO4 | ||||
Coefficient of Variation | 10% (for 10 to 100 ppm), 5% (for 100 to 300 ppm) | ||||
Shelf Life | 2 Years | ||||
Corrections for temperature & humidity | Temperature correction is necessary | ||||
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. | |||||
Possible coexisting substances and their interferences (NOTE) | |||||
Substance | Concentration | Interference | Change colour by itself to | ||
Ammonia |
| - | No | ||
Acrolein | 3/5 times | + | Red | ||
Acetone | 2 times | + | Red | ||
Methyl ethyl ketone | 6 times | + | Red | ||
MIBK | 10 times | + | Red | ||
Other substance measurable with this detector tube | |||||
Substance | Correction Factor | No.of pump strokes | Measuring Range | ||
Diacetyl | by scale | 3 | 25 to 1500 ppm | ||
Calibration gas generation Diffusion tube method | |||||
TLV-TWA | TLV-STEL | Explosive range |
| ||
- | C 25 ppm | 4 to 60% |
|
Acetaldehyde No.92
CH3CHO
Performance | |||||
Measuring Range | 5 to10 ppm | 10 to 300 ppm | 300 to 750 ppm | ||
Number of Pump Strokes | 4 | 2 | 1 | ||
Correction Factor | 0.5 | 1 | 2.5 | ||
Sampling Time | 2 minutes per pump stroke | ||||
Detecting Limit | 2 ppm (n=4) | ||||
Colour Change | Yellow → Red | ||||
Reaction Principle | 3CH3CHO + (NH2OH)3H3PO4 → H3PO4 | ||||
Coefficient of Variation | 10% (for 10 to 100 ppm), 5% (for 100 to 300 ppm) | ||||
Shelf Life | 2 Years | ||||
Corrections for temperature & humidity | Temperature correction is necessary | ||||
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. | |||||
Possible coexisting substances and their interferences (NOTE) | |||||
Substance | Concentration | Interference | Change colour by itself to | ||
Ammonia |
| - | No | ||
Acrolein | 3/5 times | + | Red | ||
Acetone | 2 times | + | Red | ||
Methyl ethyl ketone | 6 times | + | Red | ||
MIBK | 10 times | + | Red | ||
Other substance measurable with this detector tube | |||||
Substance | Correction Factor | No.of pump strokes | Measuring Range | ||
Diacetyl | by scale | 3 | 25 to 1500 ppm | ||
Calibration gas generation Diffusion tube method | |||||
TLV-TWA | TLV-STEL | Explosive range |
| ||
- | C 25 ppm | 4 to 60% |
|