硫醇类检测管70 Mercaptans R-SH
Mercaptans No.70
R-SH
| Performance | 
| Measuring Range | 0.5 to 5 ppm | 5 to 120 ppm | 
| Number of Pump Strokes | 2 to 10 | 1 | 
| Correction Factor | 1/2 to 1/10 | 1 | 
| Sampling Time | 2 minutes per pumpstroke | |
| Detecting Limit | 0.2 ppm (n=10) | |
| Color Change | White → Yellow | |
| Reaction Principle | 2R | |
| Coefficient ofVariation | 10% (for 5 to 40ppm), 5% (for 40 to 120 ppm) | |
| ShelfLife | 3Years | |
| Correctionsfor temperature & humidity | Temperature correction is necessary | |
| Store the tubes at cool and darkplace. | ||
| Possible coexisting substances and theirinterferences | (NOTE) | 
| Substance | Concentration | Interference | Change color by itself | 
| Carbonmonoxide | 2000 ppm | Plus error | Pale brown for wholelayer | 
| Hydrogensulfide | 500 ppm | Plus error | Nodiscoloration | 
| Acetylene,Ethylene | 2000 ppm | Plus error | Pale brown for wholelayer | 
| Othersubstance measurable with this detector tube | 
| Substance | Correction Factor | Pump Strokes | Measuring Range | 
| Eutylmercaptan | 1.0 | 1 or 2 to 10 | 0.5 to 120 ppm | 
| Isopropylmercaptan | 2.0 | 1 | 10 to 240 ppm | 
| Metylmercaptan | 1.0 | 1 or 2 to 10 | 0.5 to 120 ppm | 
| Propymercaptan | 4.5 | 1 | 22.5 to 540ppm | 
| Calibration gasgeneration | Permeation tubemethod | |
| TLV-TWA | TLV-STEL | Explosive range | 
| - | - | - | 
 
    
 SH+ PdSO
SH+ PdSO