三氯乙烯气体检测管132HA
Trichloroethylene   No.132HA
Cl2C:CHCl
| Performance | 
| Measuring Range | 20 to 50 ppm | 50 to 500 ppm | 500 to 1300ppm | 
| Number of PumpStrokes | 2 | 1 | 1/2 | 
| CorrectionFactor | 0.4 | 1 | 2.6 | 
| Sampling Time | 45 seconds per pumpstroke | ||
| Detecting Limit | 4 ppm (n=2) | ||
| Color Change | Yellow → ReddishPurple | ||
| Reaction Principle | Cl2C:CHCl + PbO2 +H2SO4 →HCl | ||
| Coefficient ofVariation | 10% (for 50 to 100ppm), 5% (for 100 to 500 ppm) | ||
| ShelfLife | 2Years | ||
| Correctionsfor temperature & humidity | Temperature correction is necessary | ||
| Store the tubes inthe refrigerator to keep at 10oC (50oF) orbelow. | |||
| Possible coexisting substances and theirinterferences | (NOTE) | 
| Substance | Concentration | Interference | Change color byitself | 
| Nitric Oxide,Nitrogen dioxide | - | No effect | Nodiscoloration | 
| Hydrogen chloride,Chlorine, Bromine | - | Plus error | Produce reddishpurple stain | 
| Acetone | 200 ppm | No effect | Nodiscoloration | 
| 1,1,1-Trichloroethane | - | Plus error | Produce reddishpurple at 3000 ppm | 
| UnsaturatedHalogenated HCs | - | Plus error | Produces reddishpurple stain | 
| Aromatichydrocarbons | 100ppm | Minus error | Nodiscoloration | 
| Othersubstance measurable with this detector tube | 
| Substance | Correction Factor | Pump Strokes | Measuring Range | 
| 1,2-Dichloroethylene | 1.6 | 1 | 80 to 800 ppm | 
| 1,3-Dichloropropene | 0.9 | 2 | 45 to 450ppm | 
| Calibration gasgeneration | Diffusion tubemethod | |
| TLV-TWA | TLV-STEL | Explosive range | 
| 50ppm | 100ppm | - | 
 
    