            苯气体检测管121SL C6H6 Benzene,Measuring Range 1 to 20 ppm 20 to 100 ppm Number of Pump Strokes 5 1 Correction Factor 1 5 Sampling Time 2 minutes per pump stroke Detecting Limit 0.25 ppm (n=5)苯气体检测管价格
        
    
            详情介绍:
        
        苯气体检测管121SL C6H6 Benzene
Benzene No.121SL
C6H6
| Performance | 
| MeasuringRange | 1 to 20 ppm | 20 to 100 ppm | 
| Number of PumpStrokes | 5 | 1 | 
| CorrectionFactor | 1 | 5 | 
| Sampling Time | 2 minutes per pumpstroke | |
| Detecting Limit | 0.25 ppm (n=5) | |
| Color Change | White → Brownishgray | |
| Reaction Principle | C6H6 + I2O5 + H2S2O7 → I2 | |
| Coefficient ofVariation | 20% (for1 to 5 ppm), 15% (for 5 to 20 ppm) | |
| ShelfLife | 3Years | |
| Correctionsfor temperature & humidity | Unnecessary | |
| Store the tubes at cool and darkplace. | ||
| Possible coexisting substances and theirinterferences | (NOTE) | 
| Substance | Concentration | Interference | Change color by itself | 
| Hexane | 50 ppm | No effect | No discoloration | 
| Toluene | 100 ppm | No effect | No discoloration | 
| Xylene | 150 ppm | No effect | No discoloration | 
| Aromatichydrocarbons other than benzene are trapped in the brown layer inthe pretreatment tube.If the pretreatment reagent is entirelyconsumed (whole brown layer turns to dark brown),ahigher reading will be given. | 
| Calibration gasgeneration | Diffusion tubemethod | |
| TLV-TWA | TLV-STEL | Explosive range | 
| 0.5ppm | 2.5ppm | 1.3 to 7.1% | 
 
    