产品分类
PRODUCT CENTER
销售电话:时经理
18121126516
021-64835787
甲醛气体检测管91L Formaldehyde HCHO
Formaldehyde             No.91L
HCHO
| Performance | 
| MeasuringRange | 0.1 to 5 ppm | 5 to 40 ppm | 
| Number of PumpStrokes | 5 | 1 | 
| CorrectionFactor | 1 | 8 | 
| SamplingTime | 1.5 minutes per pumpstroke | |
| DetectingLimit | 0.05 ppm (n=5) | |
| ColorChange | Yellow → ReddishBrown | |
| ReactionPrinciple | 3HCHO +(NH2OH)3H3PO4 → H3PO4 | |
| Coefficient ofVariation | 10% (for 0.1 to 0.5ppm), 5% (for 0.5 to 5 ppm) | |
| ShelfLife | 3Years | |
| Correctionsfor temperature & humidity | Temperature correction is necessary | |
| Store the tubes inthe refrigerator to keep at 10oC (50oF) orbelow. | ||
| Possible coexisting substances and theirinterferences | (NOTE) | 
| Substance | Concentration | Interference | Change color by itself | 
| Aldehydes | - | Plus error | Produce reddish brownstain | 
| Ketones | - | Plus error | Produce reddish brownstain | 
| Acid Gases | - | Plus error | Produce redstain | 
| Organic acids | - | No effect | No stain bythemselves | 
| Othersubstance measurable with this detector tube | 
| Substance | Correction Factor | No. of Pump strokes | Measuring range | |||||||
| Diisobutylketone | 5.8 | 4 | 0.58 to 29 ppm | |||||||
| Methaldehyde | 0.65 | 3 | 0.065 to3.25ppm | |||||||
| Propionaldehyde | 7.6 | 1 | 0.76 to 38ppm | |||||||
| 91L Tube Reading(n=1) | 0.2 | 0.5 | 1 | 2 | 3 | 4 | 5 | |||
| Benzaldehyde(ppm) | 2 | 4 | 9 | 22 | 40 | 63 | 92 | |||
| 91L Tube Reading(n=1/2) | 0.2 | 0.5 | 1 | 2 | 3 | 4 | 5 | 
| Cyclohexanone(ppm) | 10 | 30 | 60 | 130 | 220 | 330 | 470 | 
| Calibration gasgeneration | Diffusion tubemethod | |
| TLV-TWA | TLV-STEL | Explosive range | 
| - | C 0.3 ppm | 7.0 to 73% | 
 
    