产品分类
PRODUCT CENTER

联系我们

销售电话:时经理

                18121126516

                021-64835787

  

产品详情
  • 产品名称:丙烯腈气体检测管191L Acrylonitrile

  • 产品型号:191L
  • 产品厂商:日本GASTEC
  • 产品文档:
你添加了1件商品 查看购物车
简单介绍:
丙烯腈气体检测管191L Acrylonitrile CH2:CHCN,Measuring Range 0.1 to 0.2ppm 0.2 to 6.0 ppm 6.0 to 18.0 ppm Number of Pump Stroke 4 2 1 Correction Factor,丙烯腈气体检测管价格
详情介绍:

丙烯腈气体检测管191L Acrylonitrile
Acrylonitrile                  No.191L

CH2:CHCN

丙烯腈气体检测管191L Acrylonitrile
Performance
 

MeasuringRange

0.1 to 0.2ppm

0.2 to 6.0 ppm

6.0 to 18.0ppm

Number of PumpStroke

4

2

1

CorrectionFactor

1/2

1

3

Sampling Time

3 minutes per pumpstroke

Detecting Limit

0.05 ppm (n=4)

Color Change

Yellow → Pink

Reaction Principle

Pretreatmenttube: CH2:CHCN + Cr6+ +H2SO4 →HCN
Detector tube : 2HCN +HgCl2 → 2HCl + Hg(CN)2
HCl + Base → Chloride

Coefficient ofVariation

10% (for 0.25 to 1ppm), 5% (for 1 to 6 ppm)

ShelfLife3Years
Correctionsfor
temperature & humidity
Unnecessary
Store the tubes at cool and darkplace.

Possible coexisting substances and theirinterferences(NOTE)


Substance

Concentration

Interference

Change color by itself

Hydrogenchloride

-

No effect

Nodiscoloration

Hydrogencyanide

-

No effect

Nodiscoloration

Nitriles(丙烯腈气体检测管191L AcrylonitrileC3)

-

Plus error

Discolor pinkstain

Acetonecyanohidrin

-

Plus error

Discolor pinkstain

Alcohols, Esters,Ketones

-

No effect

Nodiscoloration

Aromatichydrocarbons

-

No effect

Nodiscoloration



Othersubstance measurable with this detector tube


Substance

CorrectionFactor

No. of PumpStrokes

MeasuringRange

n-Butyronitrile

30

1

6 to 180 ppm

2-Methyl-3-butenenitrile

2.0

2

0.4 to 12 ppm

2-Pentenenitrile

1.2

2

0.24 to 7.2ppm

3-Pentenenitrile

2.0

2

0.4 to 12ppm



Calibration gasgenerationDiffusion tubemethod



TLV-TWATLV-STELExplosive range
2ppm-3 to17%