丙烯腈气体检测管191 Acrylonitrile
Acrylonitrile                  No.191
CH2:CHCN

| Performance | 
| MeasuringRange | 2 to 5 ppm | 5 to 120 ppm | 120 to 360 ppm | 
| Number of PumpStroke | 4 | 2 | 1 | 
| CorrectionFactor | 0.4 | 1 | 3 | 
| Sampling Time | 1 minute per pumpstroke | ||
| Detecting Limit | 1 ppm (n=4) | ||
| Color Change | Yellow → Red | ||
| Reaction Principle | Pretreatment tube:CH2:CHCN + Cr6+ +H2SO4 →HCN | ||
| Coefficient ofVariation | 10% (for5 to 40 ppm), 5% (for 40 to 120 ppm) | ||
| ShelfLife | 3Years | ||
| Correctionsfor temperature & humidity | Temperature correction is necessary | ||
| Store the tubes at cool and darkplace. | |||
| Possible coexisting substances and theirinterferences | (NOTE) | 
| Substance | Concentration | Interference | Change color by itself | 
| Nitriles (C3) | 10ppm | Plus error | Discolors redstain | 
| Acetonecyanohidrin | 10ppm | Plus error | Discolors redstain | 
| Alcohols, Esters,Ketones | - | - | Nodiscoloration | 
| Aromatichydrocarbons | - | - | Nodiscoloration | 
| Hydrogenchloride, | - | - | Nodiscoloration | 
| Othersubstance measurable with this detector tube | 
| Substance | Correction Factor | Pump Strokes | Measuring Range | 
| Propionnitrile | 1.0 | 4 | 50 to120% | 
| Calibration gasgeneration | Diffusion tubemethod | |
| TLV-TWA | TLV-STEL | Explosive range | 
| 2ppm | - | 3 to17% | 